6.6 Initial installation
1. Connect and check the raw water connection (1),
permeate flow (2), permeate back flow (3) and waste
water (4)
2. Open the inlet valve manually. To do this, turn the white
magnet coil 45° counter clockwise.
Now water is running into the device.
3. Carefully open the screw for the pump venting until a
water outlet can be seen. Then close the screw again and
reset the solenoid.
4. Make sure that the wall-mounted taps at the start and end
of the ring are open.
5.
6.
Start the device using the toggle switch.
Check all connections for leaks.
7.
8.
Open the shut-off valve permeate to the drain.
Allow the device to discard permeate for at least 30
minutes.