6. Installation
The installation must be conducted by the manufacturer or by personnel trained
and authorized by the manufacturer.
6.1 Environmental Condition
Conditions for the osmosis room:
Relative air moisture < 90% at 20°C
Room temperature between +10°C and +35°C (frost protected)
Equipped with floor drain, water supply and electrical supply
6.2 Assembly
Bring the device into the appropriate position
Adjust machine feet until the device stands level and secure on the floor.
Do not store easily flammable materials in the vicinity of the device.
Do not store chemicals in the vicinity of the device.
Only operate the device with the necessary water pre-treatment.
Room of osmosis may not be freely accessible. (Access for instructed personnel
only)