Fehler! Verwenden Sie die Registerkarte 'Start', um Title;Title dem Text zuzuweisen, der hier angezeigt werden soll. |
12 | 04.10.2021 | Page 22
7.4 Emergency operation
Only use emergency operation if the automatic function fails. Have device
repaired as soon as possible.
Attention!
There is no monitoring of the water inflow during the emergency operation.
Therefore a continuous water inflow has to be guaranteed.
Absent water causes the destruction of the pump.
1. Open the inlet valve manually. To do this, turn the white
magnet coil 45° counter clockwise.
Now water runs into the device.
2. Set the key switch to position 1. Pump starts up
To turn off the device, turn switch to position 0.
In emergency operation all automatic functions are turned off. No cleaning cycle
and no automatic start and/or stop will be conducted.