75
Read/write registers
Figure 44 Setting the auto-start function
Setting the
emission delay
To set the emission delay timer use register 0x3A. Input the time in seconds into
the register.
Figure 45 Setting the emission delay
Narrow and wide
range setting
To alternate between Wide and Narrow range setting, set bit 3 of register 0x31,
where ‘0’ sets the laser to operate in Wide range and ‘1’ sets the laser to operate
in Narrow range. In the example shown in Figure 46, bit 3 is set as 1.

Table of Contents