E Series Automated Dispensing System
70 www.nordsonefd.com info@nordsonefd.com 800-556-3484 Sales and service of Nordson EFD dispensing systems are available worldwide.
Calibrating the Tip Height
Tip height is the distance from the tip to the xture plate surface. The tip height must be calibrated and then
recalibrated as needed to compensate for slight variations in height that occur when changes are made to the
system, primarily nozzle or tip change-out.
NOTE: For information on when to calibrate the tip height, to “About Tip Height” on page19.
Systems without a Tip Detector
If your system does not include the optional tip detector, follow these procedures to calibrate the tip height and then
to manually recalibrate the tip height after a same-to-same dispensing tip change.
NOTE: The optional tip detector can be added to a existing system. Refer to “Tip Alignment Kit” on page79.
Set a Calibration Point (Initial Setup for Needle Adjust)
The system uses a calibration point for the Needle Adjust function to recalibrate the tip height after a same-to-same
dispensing tip change.
PREREQUISITES
The system is in the Teach Mode. Refer to “How to Switch from Run Mode to Teach Mode” on page44.
# Key Press Step Teach Pendant Display
1
> >
• Press SETUP.
• MOVE UP / DOWN to CALIBRATION
POINT.
• Press ENTER.
[SETUP] 1/3
01 XY Move Speed
02 Z Move Speed
03 Point to Point Arc Jump
04 Park Position
05*Calibration Point
06 Mark Points
07 Program Output Status
2
>
• Jog the tip down until it is as close to
the xture plate surface as possible.
• Press F1 to save the setting.
Calibration Point
------------------------------
Jog to position
------------------------------
[F1] OK