3.3.7.1 Midlife Electronics
Midlife electronics change the power levels with a simple command in the terminal emulator:
Voltage level
Command
Max current
RAW
OPWRBA
500 mA
+5V
OPWR5V
250 mA
+12V
OPWR12
100 mA
3.3.7.2 Pre-midlife Electronics
You can change the output voltage level that is supplied to external sensors. This is typically needed
if an external analog sensor (e.g. OBS) operates in a specific voltage range and the raw battery
power does not fall within this range. The default output voltage level is RAW (battery or external
adapter). This can be changed to +5V or to +12V by changing the position of a 0-W resistor. The
three resistor positions share one pad so the modification should be straightforward.
Voltage level
Resistor position
Max current
RAW
R270 (default)
500 mA
+5V
R268
200 mA
+12V
R269
50 mA