9600 Periodic Maintenance Procedure 00-880606-02
Page 22
HANDSWITCH CONTROL
1. Connect the handswitch control to the C-Arm power panel (if it isn’t already).
2. Press the SCOUT FLUORO handswitch and verify that the X-ray ON switch illuminates and an
audible alarm can be heard as long as the switch is pressed. Verify that the averaging LEDs turn off
on the Workstation front panel while in Scout Fluoro.
3. Press the SAVE handswitch. Verify a number (1-200) is displayed in the center of the right monitor.
4. Press the PROC/BOOST handswitch and verify that the X-ray ON switch illuminates and an
audible alarm can be heard as long as the switch is pressed.
5. Enable BOOST on the control panel, press the PROC/BOOST handswitch for 5 seconds and verify
the alarm sounds at twice the normal rate.
CONTROL PANEL CONTROLS
CONTROL PANEL DIAGNOSTICS
1. Press and hold the C-ARM MODE button until the message "Loading Status" appears on the
control panel display.
NOTE: Eventually the message "STATUS V n.nn" will indicate that you have entered the status mode.
NOTE: Scroll through the choices listed within the menu by pressing either the kVp or mA switches.
2. Scroll to the Panel Diagnostics menu and select it by pressing any switch on the control panel.
3. Run the Panel LED Test, Panel Key Test, and the Panel kV/mA Test.
NOTE: To read the information or values contained under a menu item, display the item name and
press any control panel switch.
4. Select Exit This Menu and press any control panel switch.
5. Select Exit To System and press any control panel switch.
NOTE: While returning to normal system operation, the Control Panel Display will display the
message: "Loading System."
NORMAL/MAG1/MAG2
1. Press the X-ray field size key and verify that it can select each of the three field sizes. Each field
size’s LED should light when selected.
L-R
1. Press the L-R key and verify that it swaps video images on the left and right Workstation monitors.
CART MODE
1. Press the CART MODE key repetitively and verify that the Workstation toggles through all cart
modes (dependent on purchased options).