9600 Periodic Maintenance Procedure 00-880606-02
Page 31
5. Continue the exposure while rotating the camera. Verify that the new image is subtracted from the
original mask.
ZOOM (OPTION)
Verify proper operation of Zoom with the following steps:
1. Display an image on the left monitor.
2. Press the ZOOM key on the trackpad panel and observe the region of interest (ROI) box on the left
monitor displays 2X magnification.
3. Press the SUPER KEY to cycle through 2X and 4X magnification.
4. Press the TRACKPAD to move the ROI box within the image.
SHARPEN (OPTION)
Verify proper operation of Sharpen with the following steps:
1. Take an exposure and press the SHARPEN key on the trackpad panel.
2. Verify that the dark image detail is enhanced.
3. Take another exposure and then press and hold SHARPEN while pressing the trackpad.
4. Verify the amount of edge enhancement varies.
DIGITAL FLUOROGRAPHY (PULSE)
1. Cover the II with a lead cover, then press the BOOST ENABLE switch for 6 seconds to enable the
boost function.
2. Select the MAG1 field size.
3. Press the DIGITAL PROCESS FLUORO footswitch and verify the kVp tracks to 120 kVp.
4. Press the PROC BOOST handswitch and verify the mA remains at 12 mA (standard systems) or
60 mA (Cardiac systems) and the audible alarm beeps at twice the normal rate.
CARDIAC (OPTION)
Verify proper operation of the following functions on a Cardiac system:
1. If the system is Cardiac-only, the Cardiac ABS table is automatically loaded during bootup.
2. If the system is Cardiac-only, pressing the ABS select key displays the word CARDIAC on the C-
Arm control panel.
3. The Cardiac autohistogram is automatically selected.