EN-16 MB90
Press Lock out to access the lock out menu.
Press Lock to lock out
Press Unlock to release.
6.2 Calibration
6.2.1 Weight calibration
The Moisture Analyzer can be calibrated with an external mass of 50 grams. Calibration of
the Moisture Analyzer is not absolutely necessary for a correct moisture determination as the
measurement is relative. The balance determines the weight of the sample before and after
drying and the moisture is calculated on the basis of the ratio between wet and dry weights.
Nevertheless, you should calibrate the built-in balance under the following conditions:
- If this is stipulated by your quality assurance system (GLP, GMP, ISO 9001).
- If you suspect the analyzer has been abused.
To calibrate the analyzer, processed as follows:
Press Setup button to access the Setup
menu.
Press Back to return to Home screen.