MB90 EN-43
Weight calibration printout
Temperature calibration printout
WEIGHT ADJUST EXTERNAL
1.Jan 2016 11:20
Halogen Moisture Analyzer
Type MB90
SNR(Drying Unit) 1234567
SNR(Terminal)
Weight ID
Nominal Weight
50.000 g
Actual Weight
50.000 g
Difference 0.000 g
Cell Temperature 24.35 °C
Adjustment Done
Signature:
..........................................
------------END---------------
TEMPERATURE ADJUSTMENT
1.Jan.2016 11:25
Halogen Moisture Analyzer
Type MB90
SNR(Drying Unit) 1234567
SNR(Terminal)
Temp Kit ID
Temp1 target 100 °C
Temp1 actual 99 °C
Temp2 target 160 °C
Temp2 actual 161 °C
Adjustment Done
Signature:
...........................................
------------END----------------
Test data printout
Statistics printout
Interval at 30 seconds.
MOISTURE DETERMINATION
Halogen Moisture Analyzer
Type MB90
SNR(Drying Unit) 1234567
SNR(Terminal)
SW(Drying Unit) 1.20
SW(Terminal) 1.01.01
Method Name Method 1
Drying Prog Standard
Drying Temp 105°C
Switch Off A60(1mg/60s)
Start Weight 3.098 g
00:00 min 0.00%MC
--- STATISTICS DATA ---
Halogen Moisture Analyzer
Type MB90
SNR(Drying Unit) 1234567
SNR(Terminal)
SW(Drying Unit) 1.20
SW(Terminal) 1.01.01
Sample Number 6
Last Data 86.23 %MC
Mean Value 87.01 %MC
Standard Deviation 2.06
Minimum Value 83.47 %MC
Maximum Value 88.98 %MC