4. Maintenance
Access
Successively press the keys Menus and Maintenance.
1. Line On/Off
Sets the line to stop (the line is not powered and the detectors are at stop; no
event can be generated from then on).
2. Detector On/Off
Sets the detector to stop (no event can be generated from then on) if it was
not issuing an alarm or failure.
3. Test On/Off
This allows for the verification of the proper operation of a detector. In this
mode, recordings and alarm relays are suppressed.
4. Simulation
Upon its selection, the message “The controller no longer ensures detection”
is displayed.
The controller no longer keeps account of inputs (detectors, logic inputs).
The simulation measurements/status are initialized to the current
measurement/status values. The relays, the internal buzzer, and the
analog outputs remain in their current status.
The screens, management of relays, outputs, etc. are those of Normal
operation.
The internal relay and the common default LED are activated.
To change the value of a detector, use the  keys to increase or
decrease the measurement value simulated from 15% to 115%. For a
logic input, use the  keys to select the input, to select Alarm or
Alarm Off.
The banner of alarms does not appear.
The events log indicates Begin Simulation and End Simulation.
Exit the simulation mode by pressing the End simul key. Automatic
release then occurs and resets the average values to zero. The current
measurements are displayed once more.