Loading...
Test Programs Maintenance Manual
3-10
Page
No. 5
Macro table
vertical
movement
check
<Macro table vertical movement check>
This program can be used to check the vertical movement of the Top Macro inspection table.
Press the [Start] button.
Select the operation speed using the menu buttons on the liquid crystal panel.
[M1]: High speed <Fast speed> [M2]: Medium speed <Middle speed> [M3]: Low speed <Slow
speed>
If you do not specify anything here, the loader will operate at the high speed [M1]: High
speed<Fast speed> setting.
Press the [Start] button and the loader starts the operation.
If the loader has not been initialized, initialization starts.
Raises the macro table Lowers the macro table
The loader then repeats steps to .
Press the [Start] button to pause the operation at the break point.
The speed can be changed during the pause.
When an error occurs, the loader displays the same error code as in the regular operation, and then
stops.
Press the [Quit] button, and the macro table returns to its initial position and the loader stops the
operation.
Do not continue the operation for longer than 5 minutes.
Wait for longer than one minute before restarting the operation.
No. 6
L-arm
vertical
movement
check
<L-arm vertical movement check>
This program can be used to check the L-arm vertical movement.
Press the [Start] button.
Select the operation speed using the menu buttons on the liquid crystal panel.
[M1]: High speed <Fast speed> [M2]: Medium speed <Middle speed> [M3]: Low speed <Slow
speed>
If you do not specify anything here, the loader will operate at the high speed [M1]: High speed
<Fast speed> setting.
Press the [Start] button and the loader starts the operation.
If the loader has not been initialized, initialization starts.
Raises the L-arm (Lower limit Middle point) Raises the L-arm (Middle point → Upper limit)
Lowers the L-arm (Upper limit → Middle point) Lowers the L-arm (Middle point → Lower limit)
The loader then repeats steps to .
Press the [Start] button to pause the operation at the break point.
The speed can be changed during the pause.
When an error occurs, the loader displays the same error code as in the regular operation, and then
stops.
Press the [Quit] button, and the L-arm returns to its initial position and the loader stops the operation.
Do not continue the operation for longer than 5 minutes.
Wait for longer than one minute before restarting the operation.
CAUTION