Maintenance Manual Test Programs
3-13
Page
[M2]Move continuously at the same height as that in normal transfer after the F-arm is raised <Normal
height>
Press the [Start] button.
Select the operation speed using the menu buttons on the liquid crystal panel.
[M1]: High speed <Fast speed> [M2]: Medium speed <Middle speed>
[M3]: Low speed <Slow speed>
If you do not specify anything here, the loader will operate at the high speed [M1]: High speed
<Fast speed> setting.
Press the [Start] button, and the loader starts the operation.
If the loader has not been initialized, initialization starts.
Move the F-arm horizontally (Horizontal middle point position → Macro position)
Raises the F-arm (Lower limit → Upper limit)
Move the F-arm horizontally (Macro position → Stage position)
Lowers the F-arm (Upper limit → Middle point)
Move the F-arm horizontally (Stage position → Horizontal middle point position)
Lowers the F-arm (Middle point → Lower limit)
The loader then repeats steps to .
Press the [Start] button to pause the operation at the break point.
The speed can be changed during the pause.
When an error occurs, the loader displays the same error code as in the regular operation, and then
stops.
The above operation pattern indicates the F1-arm (upper side) operation.
Press the [Quit] button, and the F-arm returns to its initial position and the loader stops the
operation.
No. 10
Centering
sensor count
check
<Centering sensor count check>
This program can be used to check the centering sensor count state.
Press the [Start] button.
Initialized.
Press the interlock switch and the [Start] button.
FOUP Door open.
When the door opens completely, release your hand from the interlock switch and press the [Start]
button. ( to are for AL120-LMB12-LP3 only.
Raises the A-arm and turns the horizontal drive motor OFF.
Set a wafer on the A-arm so that it is in virtually the same position as during transfer.
Press the [M2] button <Vacuum valve change> to hold the wafer.
Press the [M3] button<Search>, and press the[START] button.
The A-arm moves horizontally and the value detected by the centering sensor is displayed on the
liquid crystal panel.
The following four items are displayed:
S1:#1 Displacement of the detected position toward the front of the sensor [um],
displacement of the detected position toward the back of the sensor [um]
S2:#2 Displacement of the detected position toward the front of the sensor [um],
displacement of the detected position toward the back of the sensor [um]
S3:#3 Displacement of the detected position toward the front of the sensor [um],
displacement of the detected position toward the back of the sensor [um]
S4:#4 Displacement of the detected position toward the front of the sensor [um],
displacement of the detected position toward the back of the sensor [um]
#3 and #4 are displayed only on the AL120-6, not displayed on other models.