Test Programs Maintenance Manual
3-22
Page
No. 15
Notch
alignment
adjustment
<Notch alignment adjustment>
This program can be used to check and adjust the notch positions f.
Insert a wafer into any slot of a cassette. and place the cassette on the cassette installation part.
Using the 4-way button, select the number of wafer type to adjust or register from the list of wafer
type numbers displayed on the liquid crystal panel, and then press the [Start] button.
(Registration is done for each wafer diameter. Registration for each wafer thickness is not
available.)
Mapping is performed and the [Wafer No.] button for a slot in which there is a wafer blinks.
Press the [Wafer No.] button to specify the slot into which the wafer to be adjusted has been
inserted. The [Wafer No.] button for the selected slot stops blinking and stays lit.
Press the [Start] button.
The selected wafer is loaded from the cassette.
Adjust the notch positions.
Data saved in the internal memory is displayed on the liquid crystal panel.
<CW/CCW 00>
Set the notch position setting knob in the direction you want to adjust the notch alignment.
Correct the notch alignment position. Save the correction data in the internal memory.
Use the 4-way button to correct the position.
4-way button [RIGHT]: Clockwise
4-way button [LEFT]: Counterclockwise
One pulse: About 0.1 degree
Register the correction value.
[M1]: Register <Save>
Press the [Start] button to repeat steps to .
If you do not press [M1] button but press the [Quit] button instead, the wafer is unloaded into the
cassette without saving the data.
Use the [Wafer alignment] control for adjusting the orientation flat positions.
Correction data must be registered separately for each direction of notch.
[M3] Unload
[M4] Load
No. 16
L-arm
rotation
origin
sensor
position
check
< L-arm rotation origin sensor check >
This program can be used to check and adjust the L-arm rotation origin.
Press the [Start] button.
Select the item to adjust using the menu buttons on the liquid crystal panel.
[M1]Check the L-arm rotation origin sensor <L-arm rotation check>
[M2]Check the L-arm wafer support unit <L-arm wafer support unit check>