b)
By using a screwdriver
or
tweezers, adjust the
casing
to
the center
of
the inner guide
unit.
(See
Fig. 4-41)
Fig.441
c)
By using Torque screwdriver
OT0044,
tighten
two
Setscrews GUK2.6x10SA
for
the
fixing
piece.
(See
Fig. 4-42)
Tightening torque:
800
g-cm
d)
Apply
Shellac OT1131
to
the screws to
prevent loosening.
e)
Finally
fix
inner guide
AA819700
by
tighten-
ing three Screws AB3x10SA.
Fig.
442
5.
DISASSEMBLING
PROCEDURES FOR
C-GH
CONDENSER
HOLDER
5-1
By removing
two
Screws
GUK3x10SA,
dis-
mount
left
Dovetail
AA870800.
(See
Fig. 5-1)
Fig. 5-1