Olympus CH Series - Disassembling and Reassembling Procedures for C-MVR; Disassembly Procedure of East-West Drive Guide

Olympus CH Series
58 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
10·2-8 Adjust the guide
a)
Tighten
two
Adjusting screws
HU3x4SA
by
using Torque screwdriver OT0044.
(See
Fig. 10-16)
Tightening force:
700
g·cm
b) Finally tighten three Screws
AB3x10SA
on
the outer
gu
ide.
NOTE.
For later steps, proceed
to
Section 5
of
this
manual.
-
-;.
Fig.
10·16
11. DISASSEMBLING
AND
REASSEMBLING PROCEDURES FOR C·MVR
J
QJ
Fig.11·1
o
AA877900
AA878000
o
3PSK2x6SA
11-1
Disassembly procedure
of
East-West Drive
guide
11·1·1 Remove Specimen holder
unit
by
drawing
4 Screws 3PSK2x6SA.
(See
Fig. 11·1)
11-1-2 Remove Rack
AA877900
by
drawing 3
Screws 3PUK2x6SA.
(See
Fig. 11-1)
11-1-3 Remove East·West Graduated plate
AA-
878000
by
drawing 3 Screws PUK2x3SA.
(See
Fig. 11-2)
----------
--------
Fig. 11·2
PUK2x3SA
-45-