Olympus CH2 Series - Sensitivity adjustment of fine focus movement

Olympus CH2 Series
41 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
8.
8-1.
B-2.
Check procedure
7-2-3. Check for equal clearance between the jig
and the stage surface with feeler gauge
(OT0317) at the front or back end. Maxi-
mum tolerance between stage and jig =
0.15 mm.
7-2-4.
If the reading exceeds tolerance, adjust the
tilt by inserting a piece of tin foil between
the stage (AA871800) and the substage
block (AB469400).
Accuracy adjustment of fine focus scale
adjustment
Assemble the microscope as indicated in Fig. 8-2-3. The difference between the two reading
Observation tube
obtained in step 8-2-l and step 8-2-2
Objective 40X No special jigs are
should be within 1.5 divisions.
Eyepiece 1 OX
required.
8-2-4. Repeat steps 8-2-l and 8-2-2 after removing
Condenser
the block (B2KCOOlO).
Specimen (Any specimen good for 40X)
Block (B2KCOOlO)
8-3. If the readings exceed the above mentioned
tolerances:
8-3-l. Recheck assembly step l-2 and check for
8-2-l. Focus on the specimen and check the fine
damages on components.
focus scale reading.
8-3-2. Recheck assembly step 3-1 and check for
8-2-2. Rotate the fine focus adjustment knob by
damages on components, especially gears.
of. 25 pm (10 divisions), then refocus on the
specimen and take the scale reading.
-41
-