7.2 Surface disinfectant cleaner
95
CLV-190 INSTRUCTION MANUAL
Ch.7
CAUTION
Do not reprocessing the AC mains power inlet. Reprocessing them can deform or
corrode the contacts, which could damage the light source.
Do not soak in water and autoclave, the light source and accessories. These
methods will damage them.
Do not wipe the external surface with hard or abrasive wiping material. The surface
will be scratched.
When patient debris enters a hole or gap of the light source, contact Olympus
without disinfecting it. If you try to disinfect by force, the disinfectant solution will get
inside the light source causing fire and malfunction of the light source.
Use a medical-grade hydrogen peroxide surface disinfectant cleaner with properties as shown in
Tab le 7 .1 .
NOTE
Use the surface disinfectant cleaner cleared/approved by your national regulatory
agency.
Table 7.1 Hydrogen peroxide surface disinfectant cleaner with properties
Follow the surface disinfectant cleaner manufacturer’s instructions regarding concentration, drying,
temperature, contact time, use life, and expiration date. The surface disinfectant cleaner shown in
Table 7.2 was used for validation.
Table 7.2 Surface disinfectant cleaner used for validation
7.2 Surface disinfectant cleaner
Surface disinfectant cleaner Hydrogen peroxide
Percentage solution Undiluted solution
Disinfectant concentration Hydrogen peroxide 1.5g/100g
Component Hydrogen peroxide, Glycolic acid
Others Acting as disinfectant and cleaner
No rinsing required
Trade name Type Manufacture
Incidin™ OxyFoam S Hydrogen Peroxide ECOLAB