Illumination Examination lamp Xenon short-arc lamp (ozone-free) 300 W
Average lamp life Approximately 500 hours of continuous use (with intermittent use,
the lamp life may vary slightly.)
Ignition method Switching regulator
Brightness
adjustment
Light-path diaphragm control
Cooling Forced-air cooling
Intensity mode Normal or high intensity
NBI observation Available
Color conversion Possible using special-purpose filter.
Emergency lamp Halogen lamp (within mirror) 12 V 35 W
Average emergency
lamp life
Approximately 500 hours
Automatic brightness
adjustment
Automatic brightness
adjustment method
Servo-diaphragm method
Automatic exposure 17 steps
Air-feeding Pump Diaphragm type pump
Pressure switching 4-level available (OFF, low, medium, high)
Water feeding Method Feeds water by pressurizing the detachable water container with
air.
Indicators on front
panel
Emergency lamp Indicates absence of emergency lamp, disconnection and use of
emergency lamp.
NBI When the NBI observation mode is enabled, the observation mode
indicator lights up.
OP It lights up if a special-purpose filter is installed into the light
source.
Setting memory Settings (except the filter setting) are stored even when the light
source is OFF.
Classification
(medical electrical
equipment)
Type of protection
against electric
shock
Class I
Degree of protection
against electric
shock of applied part
Depend on applied part. See also applied part (camera head or
videoscope).
Degree of protection
against explosion
The light source should be kept away from flammable gases.