EMC information
CLV-190 INSTRUCTION MANUAL
App.
WARNING
Portable RF communications equipment (including peripherals such as antenna
cables and external antennas) should be used no closer than 30 cm (12 inches) to
any part of this instrument, including cables specified by Olympus. Otherwise,
degradation of the performance of this equipment could result.
Test frequency
[MHz]
Band [MHz] Modulation
*1
*1 The carrier shall be modulated using a 50% duty cycle square wave signal.
Maximum power
[W]
Immunity test
level [V/m]
385 380 – 390
Pulse modulation
*1
18 Hz
1.8 27
450 430 – 470
FM ±5kHz
deviation 1 kHz sine
228
710
704 – 787
Pulse modulation
*1
217 Hz
0.2 9745
780
810
800 – 960
Pulse modulation
*1
18 Hz
228870
930
1720
1700 – 1990
Pulse modulation
*1
217 Hz
2281845
1970
2450 2400 – 2570
Pulse modulation
*1
217 Hz
228
5240
5100 – 5800
Pulse modulation
*1
217 Hz
0.2 95500
5785