45
Problem Cause Remedy Page
The picture is not in focus. The specimen is not in focus
properly.
Adjust the focus correctly with the
fine adjustment knob.
The aperture iris diaphragm of the
condenser is open too wide.
Close the aperture iris diaphragm
a little.
The field iris diaphragm is not set
properly.
Adjust the field iris diaphragm until
the image circumscribes the field
of view.
Lens components of the micro-
scope are contaminated or the
cover glass on the front of the
camera is stained.
Clean the objective, camera
adapter lens, condenser and/or
window lens of the microscope,
or clean the cover glass on the
bottom of the camera head.
14
The microscope and/or camera
are subjected to vibration during
recording.
Acquire images in an environment
in which the microscope and
camera are not vibrated. It is effec-
tive to use an anti-vibration bench.
The 4140 x 3096 and 2070 x 1548
images are not neat.
The camera is subjected to
vibration during recording.
Acquire images in an environment
in which the microscope and
camera are not vibrated. It is effec-
tive to use an anti-vibration bench.