Simplified Transmitted Polarized Light Observation
}The SZX-AN analyzer is required.
1. Loosen the analyzer clamping knob @, fit the analyzer into the front of the
objective ² and tighten the clamping knob (by placing it on the right
side of the microscope body).
2. With no specimen placed on the rotary stage, rotate the analyzer rotation
ring ³ so that the field of view is darkest (cross-Nikol adjustment).
3. Place a specimen on the rotary stage and rotate it to perform the
polarized light observation.
10-4 Stage Adapter Type 1 SZH-STAD1
This adapter has the same function as the SZX-STAD1 BX stage adapter type 1, but the usable stages with this adapter
is only the BH2-SH horizontal knob stage.
The assembly and polarizer (SZX-PO) installation procedures for this adapter are identical to those for the SZX-STAD1.
Refer to the previous section (page 28).