EasyManua.ls Logo

Olympus USPM-RUIII Series - Page 38

Olympus USPM-RUIII Series
143 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
PV6088-F4E017
38
Read the manuals carefully before performing lamp house inspections.
Inspections on a regular basis (at least once every half year and at the time of lamp
replacement) is recommended in order to ensure safe use.
Inspect the items in the list, and mark an X next to each passed item and a circle next
to each item having some problem.
If there are any circled items in the latest inspection, stop using the unit and ask the
agent to inspect the unit or to replace it with a new lighting device.
If you discover a failure in any items not on the list or a failure in any other components
of our product, other than the lighting device, stop using the unit and ask your
supplier to inspect the unit.
Repairs, replacement and inspections out of warranty are available for a fee.
Contact your supplier with any questions you may have.
Result
(date of inspection)
Inspection checklist
/
/
/
/
1. Eight years have elapsed since purchase, or the total time of lamp use has
exceeded 20,000 hours.
2. Sometimes lamp does not illuminate even when you turn on the main switch
(except for discharge-type lamps
*1
).
3. The light blinks if you move the lamp cable or the body itself.
4. The lamp cable or another part is abnormally hot.
5. There is an odor like something is burning when in use.
6. Light flickers even after the lamp is replaced (except for discharge-type lamps
*1
).
7. A deformation, unsteadiness or loose part is discovered when installing the
lighting device (e.g., a cover is too tight to open).
8. The color of any terminal or the lamp socket is drastically changed. Or, the color is
different between the left one and the right one (except for discharge-type
lamps
*1
).
9. There is change of color, deformation or cracks in the appearance of the device.
10. There is melting, cracks, deformation or solidification in the lamp cable or wiring.
11. Frequency of repair is higher than before for the same type of other units brought
into service around the same time.
* Make a copy of the list if you find some of the check result fields have already been used.
*1 Discharge-type lamps: mercury lamps, xenon lamps, and metal halide lamps
Checklist for inspection of the lighting device (lamp house)

Related product manuals