6. Setting the Wavelength Calibration Data
If you are using the USB-cable version, perform the following settings.
(1) Double-click to open the “USPM.int” file located in the folder into which the
USPM-RU/RUIII program has been installed (C:¥USPM¥ or ¥USPM440 by default). If
it cannot be opened, right-click on the “USPM.int” file, select “Open with”, and select
Notepad to open it.
(2) Enter the numerical values found on the measuring head, as shown below, and save
the file.
1: 38 2: 77
3: 214 4: 401
5: 454 6: 485
Example of the numerical
values on the measuring head
Example of the description in the USPM.int file
1st line : (Data file identification tag)
2nd line : (Configuration file name)
3rd line : (Used port number)
4th line : 38 Set the numerical value for item 1 in the
box on the left.
5th line : 77 Set the numerical value for item 2 in the
box on the left.
6th line : 214 Set the numerical value for item 3 in
the box on the left.
7th line : 401 Set the numerical value for item 4 in
the box on the left.
8th line : 454 Set the numerical value for item 5 in
the box on the left.
9th line : 485 Set the numerical value for item 6 in
the box on the left. (*1)
Last line : (Shutter waiting time)
*1: There is no 9th line in the 440-840 nm specification.