7. Details of [Standard] Menu
PV6253-F4E016 27
7. Details of [Standard] Menu
USPM-RU compares measured data to the preset standard and judges whether the data meets it or
not.
7.1. [Standard setting] menu
In the [Standard setting] menu, the standard for OK/NG judgment is set up, and the
reference values are entered. When the window is closed with the [OK] button, the program
judges whether the selected data meets the standard or not. NG is decided if the upper-limit
standard value is exceeded or the lower-limit standard value is undercut. The status bar at the
screen bottom shows the name of the standard file currently open.
7.1.1. Reflectance
When the [Enable reflectance] check box is checked, the program decides OK or NG
using the reflectance. Setting can be made in Condition 1 up to Condition 10, and the
conditions currently checked only are valid. Key in the upper-limit value and lower-limit
value. The result will be indicated as Reflectance: OK/NG in the status bar. Moreover, the
standard line will be displayed on the Main Panel.
When the [Enable] check box in [Average reflectance] is checked, the program decides OK
or NG using the mean reflectance. Key in the upper-limit value and lower-limit value. The
result will be indicated as Rm: OK/NG in the status bar.
When the [Enable] check box in [Judge: Luminous reflectance] is checked, the program