77
SAMPLING TIME: 3 samples per second. Re-zero of CJC
and auto calibrate every 5 seconds
TEMPERATURE
COEFFICIENT: <150ppm/°C of max. linearized range
(typical)
REFERENCE
CONDITIONS: 22°C ±2°C, 115/230 V ±15% after 30
minutes settling time
SENSORS
INPUT TYPES: Thermocouple: J, K, E, R, S, N, B, J DIN:
RTDs: 2-wire Pt-100, 3-wire Pt100
0 to 20 mV, 4 to 20 mV, user selectable
EXTERNAL
RESISTANCE (TC): 100 ohms maximum
COLD JUNCTION
COMPENSATION: 0.05°/°C (20:1)
BURNOUT
PROTECTION: Fault display, upscale (downscale is key
selectable)
COMMONMODE
REJECTION: 140 dB, 240 V, 50/60 Hz
NORMAL MODE
NOISE REJECTION: 60 dB, 50 Hz