EasyManuals Logo

Omega DPF20 Series User Manual

Omega DPF20 Series
40 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #7 background imageLoading...
Page #7 background image
7
DPF20
Frequency, rate, total or period counter
Instruction Manual
Digits
number of digits 6
led 7 segments led
color red or green
digit height 14 mm
Reading
maximum reading 999999
minimum reading -199999
decimal point congurable X.X.X.X.X.X.
overrange / underrange congurable to ash, reset or preset (see
secon 1.19.16)
display refresh 15 readings / second
memory retenon yes, retains reading value in case of pow-
er loss
Signals accepted NPN, PNP, Namur, pick-up, TTL, induc-
ve, mechanical, quadrature, ...
Max. Vdc at input ±30 Vdc
Input impedance 2K4 with pull-up or pull-down resistor
470K without pull resistor
Accuracy of the quartz ±0.01 %
Thermal dri 20 ppm / ºC
Excitaon voltage congurable
output voltage +18 Vdc, +15 Vdc, +9  Vdc, +5 Vdc
maximum current 70 mA
protecon yes, current limited to 70 mA
Frequencies counter modes (see Table 3)
ratemeter modes (see Table 4)
periodmeter modes (see Table 5)
Power
power ‘H’ 85 to 265 Vac/dc
power ‘L 11 to 60 Vdc and 24/48 Vac
isolaon* 2500 Ve with power ‘H’
1500 Ve with power ‘L
*tested for 60sec.
consumpon <1.5 W only meter
<4.0 W meter with opons
Conguraon 3 buons front keypad
Front protecon IP65
Output and control opons relays, analog outputs, serial communi-
caons (see secon 2)
Mechanical
mounng panel
connecons plug-in screw terminal
housing material ABS, polycarbonate (V0)
weight <150 grams
front size 96 x 48 mm (1/8 DIN)
panel cut-out 92 x 44 mm
depth from panel 91 mm (including terminals)
Temperature
operaon from 0 to +50 ºC
storage from -20 to +70 ºC
warm-up me 15 minutes
Counter Mode Frequency Secon
Counter
‘FAST’ mode acve max. 250 KHz 1.19.2
normal mode max. 9 KHz 1.19.2
Counter + inhibion max. 9 KHz 1.19.4
Counter + control A/S max. 9 KHz 1.19.5
Counter dierenal max. 9 KHz 1.19.6
Counter
quadrature
mode x1 max. 17 KHz 1.19.3
mode x2 max. 16 KHz 1.19.3
mode x4 max. 11 KHz 1.19.3
Table 3 - Maximum input frequency for counter modes
Ratemeter Mode Frequency Secon
Ratemeter
normal mode
max. 500 KHz
1.19.7
‘SLOW’ mode ac-
ve
max. 200 Hz
min. 1 mHz
1.19.7
Ratemeter
quadrature
mode x1 max. 17 KHz 1.19.8
mode x2 max. 16 KHz 1.19.8
mode x4 max. 11 KHz 1.19.8
Table 4 - Maximum and minimum input frequency for ratemeter modes
Periodmeter Mode Frequency Secon
Periodmeter
normal mode
max. 500 KHz
1.19.9
‘SLOW’ mode ac-
ve
max. 200 Hz
min. 1 mHz
(1000 sec.)
1.19.9
Table 5 - Maximum and minimum input frequency for periodmeter modes
1.12 Mechanical dimensions (mm) (in)
1.11 Technical specicaons
44
92
Panel
cut-out
16
8
75
96
48
(1.89)
(3.78)
(0.63)
(0.31)(2.95)
(1.74)
(3.63)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Omega DPF20 Series and is the answer not in the manual?

Omega DPF20 Series Specifications

General IconGeneral
BrandOmega
ModelDPF20 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals