2.2.4 RBP1 rechargeable battery pack
Designated use
Figure 2-13: Rechargeable battery pack (RBP1)
RBP1 is a rechargeable lithium-ion battery pack provided by OMICRON for operating measuring devices
without the local power supply. RBP1 can be stacked to increase the available capacity. You can
connect up to five RBP1 rechargeable battery packs to power long-time measurement setups.
RBP1 must be used and charged properly. Before using RBP1, you must read and understand the
following warning messages. If you have any question, contact the OMICRON technical support.
WARNING
Death or severe injury caused by fire and burns possible
Do not drop or crush, do not short circuit, do not open, and do not expose the battery
to high temperatures.
Do not immerse the battery in any liquids.
Do not charge the battery in flammable environment.
Do not charge it below 0 ºC/32 ºF and above 40 ºC/104 ºF and do not operate it below
–20 ºC/–4 ºF and above 55 ºC/131 ºF
Never charge the battery near flammable objects.
Do not deep discharge the battery and recharge it regularly during storage.
Dispose of the battery properly, it may explode if damaged or disposed of in fire.
Read this User Manual before charging the battery and use the specified charger only.
Follow the local ordinances and/or regulations for usage of lithium batteries.