MPD 800 User Manual
46 OMICRON
2.3.4 Disconnection
Before entering the high-voltage area, observe the following instructions:
1. Clean the fiber-optic cables. In dusty regions, use protective caps.
2. Disconnect the cables.
3. Remove the MPD 800 acquisition unit, RBP1, CPL1/CPL2 or accessories from the high-voltage area.
4. Recharge the battery pack for later use.
WARNING
Death or severe injury caused by high voltage or current possible
► Switch off the high voltage.
► Always obey the five safety rules (see 1.2.2 "Safety rules" on page 6) and follow the
detailed safety instructions.
► Make sure to observe dangerous areas.
► Always obey the internal safety instructions for working in areas with high voltage to
avoid injury.
WARNING
Death or severe injury caused by fire and burns possible
RBP1 is a rechargeable lithium-ion battery pack.
► Do not drop or crush, do not short circuit, do not open, and do not expose the battery
to high temperatures.
► Do not immerse the battery in any liquids.
► Do not charge the battery in flammable environment.
► Do not charge it below 0 ºC/32 ºF and above 40 ºC/104 ºF and do not operate it below
–20 ºC/–4 ºF and above 55 ºC/131 ºF
► Never charge the battery near flammable objects.
► Do not deep discharge the battery and recharge it regularly during storage.
► Dispose of the battery properly, it may explode if damaged or disposed of in fire.
► Read this User Manual before charging the battery and use the specified charger only.
Follow the local ordinances and/or regulations for usage of lithium batteries.