11451 Belcher Road South, Largo, FL 33773 • USA • Tel +1 (727) 447-6140 • Fax +1 (727) 442-5699 • sales@onicon.com
D-100 Dual Network Interface Installation Guide 03/18 - 1015-2 / 35094 Page 27
REGISTER
ADDRESS
DESCRIPTION
REGISTER
RANGE
DATA
RANGE
OVER
RANGE
READ/
WRITE
COMMENT
40001
Meter Operating Mode
Indicator
1 – 3
Not
applicable
Read Only
1 – indicates single mode
2 – indicates dual mode
3 – indicates bi-directional
mode
40002 Mode Status Indicator 1 – 2
Not
applicable
Read Only
1 - indicates heating mode
or forward direction
2 - indicates cooling mode
or reverse direction
40009 Volume Rate – GPM 0 – 65535 0 – 65534 65535 Read Only
40010 Volume Rate – GPM x 10 0 – 65535 0 – 65534 65535 Read Only
40011 Volume Rate – GPH 0 – 65535 0 – 65534 65535 Read Only
40012 Volume Rate – MGD 0 – 65535 0 – 65534 65535 Read Only
40013 Volume Rate – L/S 0 – 65535 0 – 65534 65535 Read Only
40014 Volume Rate – L/M 0 – 65535 0 – 65534 65535 Read Only
40015 Volume Rate – L/Hr 0 – 65535 0 – 65534 65535 Read Only
40016 Volume Rate – M³/Hr 0 – 65535 0 – 65534 65535 Read Only
40017 Volume Rate – M³/Hr x 10 0 – 65535 0 – 65534 65535 Read Only
40018 Mass Rate – Lb/Hr 0 – 65535 0 – 65534 65535 Read Only
Mass units are only
available when using
F-2200 Vortex Meter.
40019 Mass Rate – Lb/Hr x 10 0 – 65535 0 – 65534 65535 Read Only
40020 Mass Rate – kg/Hr 0 – 65535 0 – 65534 65535 Read Only
40021 Mass Rate – kg/Hr x 10 0 – 65535 0 – 65534 65535 Read Only
40022 Supply Temperature – °F 0 – 65535 0 - 655.35
Not
applicable
Read Only
Multiply by 0.01 to read
temperature to 2 decimal
places.
40023 Return Temperature – °F 0 – 65535 0 - 655.35
Not
applicable
Read Only
40024 Supply Temperature – °C
-32768
to
+32767
-327.68
to
+327.67
Not
applicable
Read Only
40025 Return Temperature – °C
-32768
to
+32767
-327.68
to
+327.67
Not
applicable
Read Only