F
-
1000 SERIES TURBINE METER CONFIGURATION UTILITY AND INSTALLATION GUIDE
ONICON Incorporated 727.447.6140 Page 10 onicon.com
“Missing” text with red background
This status appears when the USB cable to the turbine meter
is disconnected. If the cable is connected and the fault is
still present, check the Device Manager as explained in the
“Driver Installation” section, and make sure there is no error
with the port used for the turbine meter connection. Re-
install the driver if there is a fault.
If there is no error with the port in the Device Manager, close the Conguration
Utility, remove and re- install the USB cable, and then re-open the Conguration
Utility. The meter should initialize and connect normally.
3.2.2 STATUS INDICATOR
This indicator provides information about the current status of the alarms and
warnings in the “Status” tab.
“Normal” text with green background
This status indicates that the meter does not have any
warnings or alarms.
ALARM” text with ashing orange background
This status indicates that an alarm is present on the meter. If
you open the “Status” tab, you will nd a complete list of the
current alarms.
“Warning” text with ashing yellow background
This status indicates that a warning is present on the meter. If
you open the “Status” tab, you will nd a complete list of the
current warnings.
More information about the dierent warnings and their meanings can be found
in the “Status” section on pages 30-32.
3.2.3 MODE INDICATOR
The mode indicator noties the user if the meter is in
“Running” mode or “Calibration” mode. “Calibration” mode
is only used by ONICON during a factory calibration. If you
receive a meter which states the mode is “Calibration,” please
contact ONICON immediately.