F
-
1000 SERIES TURBINE METER CONFIGURATION UTILITY AND INSTALLATION GUIDE
ONICON Incorporated 727.447.6140 Page 25 onicon.com
3.3.7 REAL TIME DATA
The real time data tab displays data related to the meter’s operation. The
turbine speeds, output levels, and equivalent ow speed based on pipe size
conguration.
The information on this tab is used to diagnose a turbine which is not spinning,
and to provide a ow rate reading without having to measure an output signal
level on the meter’s wires with a multimeter.
Real-time Data tab
Average Pulses Per Second (Read Only)
This parameter reports the average turbine signal pulses per second (Hz). The
average is the mean of the top and bottom turbine signal pulses per second.
Top Pulses Per Second (Read Only)
This parameter reports the top turbine signal pulses per second (Hz).
If the bottom turbine reports a Hz value, yet the top shows none, the turbine
meter should be inspected for debris or damage.
Bottom Pulses Per Second (Read Only)
This parameter reports the bottom turbine signal pulses per second (Hz).
If the top turbine reports a Hz value, yet the top shows none, the turbine meter
should be inspected for debris or damage.
Instantaneous Input Frequency Hz (Read Only)
This frequency is an un-scaled output and requires the meter’s “Meter Factor”
from the “Meter Data” tab in order to be converted to gpm.