Appendix: ER060AUTOL setup A
The configuration is made in the following steps.
1. Enter the IO256 software, section A.1
2. Set up units (protocol: IOMan) for communication between the analyser and the
IOMan modules, section A.2.
3. Load and save the settings from the IOMan modules (inputs and outputs),
section A.3.
4. Logg the calibration cell temperature, section A.4.
5. Configuration of the calibration unit, CU004/007, section A.5.
A.1 Enter the IO256 software
1. Start the analyzer software.
2. From the root menu press [F1] Installation and then [F5] Input signal setup. The
message “No logger installed. Install logger? (Y/N)” can be read. Press [Y] and [En-
ter].
3. The available datalogger softwares are listed one at a time. When the option “Install
IO256?” is shown, press [Y] and [Enter].
A.2 Communication units using the
IOMan protocol
1. Enter the IO256 software.
2. Choose Input => Serial communication. Define a serial communication unit. Se-
lect a com port and IRQ no. corresponding to the hardware configuration of the sys-
tem. The baud rate should normally be set to 19 200 bps. Choose the IOMan
protocol.
3. Choose Input => Sources. Define an input source (subunit) of the type IOMan. If
more than 16 input channels are needed, several subunits has to be defined. (1 subu-
nit = 16 channels).
4. Choose Output => Serial communication. Define a serial communication unit.
Repeat the settings from step 2.