7.7.2
THERMOSTAT VERIFICATION
Disconnect J3 connector from the Lamps Control board. The fan should stop.
Switch the light on. Wait approximately 15 minutes; the light should switch off
automatically.
Connect J3 back to the Control bord: the light should return within 10
minutes.