8. TROUBLESHOOTING
8.1 WARNING AND ERROR MESSAGES
The ASSISTANT console delivers warning and error messages on the LCD
to inform the operator of situations which may require special attention.
A window pops up in the middle of the LCD giving full explanation of the
warning or error, if operation can continue, an OK button appears on the
window.
The operator must depress this button to recognize that the message has
been received. The pop up window will close and a small warning symbol will
then appear, as a reminder, on the warnings frame located on the right side
of the LCD. This reminder will automatically disappear when the cause of the
warning has been removed.
The table below summarizes the warning and error messages.
A
R
N
I
N
G
R
R
O
R
Err 01: Serial comunication fault
Err 02: Low input pressure
Err 03: I/A Cassette missing
Err 04: I/A cassette full.
Err 06: Vacuum level error
Err 08: Diathermy power mismatch
Err 14: Aspiration occluded
Err 16: Low Ultrasounds power
Err 19: Pressure level error
Err 23: Air enable failure
Err 24: US enable/reading failure
Err 25: US enable failure
Err 26: Diath enable/reading failure
Err 27: Diathermy enable failure
Err 28: Footswitch failure