5.Menu Interpretation
33
Stops updating trace data and display current trace for
observation..
5.6.7 [Operations]
Enter trace math related sub menu.
1) [1 ↔ 2]
Exchange the trace stock 1 data with trace stock 2 and place them
in display mode.
2) [2-DL2]
Deduct display line value in trace stock 2. This function execute
once when activated. Press [2 - DL → 2] again to execute it the
second time. When this function activated, display line will also be
activated.
3) [2 ↔ 3]
Exchange the trace stock 2 data with trace stock 3 and place them
in display mode.
4) [1→3]
Exchange the trace stock 1 data with trace stock 3 and place them
in display mode.
5) [2→3]
Exchange the trace stock 2 data with trace stock 3 and place them
in display mode.
5.7 Detector
While displaying a wider span, each pixel contains spectrum
information associated with a larger subrange. That is, several
samples may fall on one pixel. Which of the samples will be
represented by the pixel depends on the selected detector type.
Press this key to pop out the relevant menu includes [Pos Peak],
[Neg Peak], [Normal], [Sample], [RMS Avg], [Voltage Avg],
[Quasi-Peak].
Key points:
Selects an appropriate type according to the application in order
to ensure the accuracy of the measurement for your application.