Searches the minimum from the sampling data segment and
displays it at the corresponding pixel.
5.7.3 [Normal]
When noise is detected, the positive and negative peaks are
alternately displayed, otherwise only positive peaks are displayed.
5.7.4 [Sample]
Set the detector to the sampling detector mode. This mode is
usually used for video averaging and noise frequency Maker.
5.7.5 [RMS Avg]
Set the detector to the RMS Average detector mode. This mode
calculates the RMS average power of all the samples in the
sample bucket.
5.7.6 [Voltage Avg]
Set the detector to the Voltage Average detector mode. This mode
calculates the average voltage of all the samples in the sample
bucket.
5.7.7 [Quasi-Peak]
Set the detector to the Quasi-Peak detector mode. This mode is
available when EMI filter is turned on. The quasi-peak detector is a
peak detector that is weighted by the duration and repetition rate of
the signal, as specified by the CISPR 16-1-1 standard. Quasi-peak
detection is characterized by a fast charge time and slow decay
time.
5.8 【Display】
Controls the screen display of the analyzer, such as full screen,
setting the on or off for window zoom, display line, amplitude scale,
grid and label.
5.8.1 [Full Screen]
Set to full-screen display graphical interface, press any key to exit.