Pop up system information and system log soft menu.
1) [System Info]
2) [Firmware Update]
3) [Option]
5.19.1.1[System Info]
Press to display system information.
5.19.1.2 [Firmware Update]
1. Create a folder named “spectrum” (lowercase) on the root
directory of the USB memory device, and copy the firmware file
onto this folder.
2. Insert the USB memory device into the front-panel USB
connector on your instrument. Press 【System】key on the front
panel, press [System ►], and press [Firmware Update] to
execute firmware update.
3. The analyzer will perform the update process. The upgrade
procedure will take approximately 30 seconds. During the
update process, do not remove the USB memory device, do not
power off the instrument or press any key. If the update process
fails, please report the problem to your distributor or our
technical support.
4.Once the upgrade is completed, the instrument will automatically
restart
5.19.1.3 [Option]
Enter option TG or EMI function configuration.
5.19.1.4 [Return]
Return to the previous menu.
5.19.2 [Setting]
A soft menu for setting the interface address of the spectrum
analyzer, including [Network ►]. The spectrum analyzer supports
VGA, LAN and USB interface communication.