© 2024 PacBio. All rights reserved. For research use only. Not for use in diagnostic procedures.
PN 103-140-400 REV05 JUN2024
With the clustering reagent plate still housed in the reagent drawer, use a fresh serological pipette tip to add
25 mL of the 0.1% hypochlorite solution into the previously punctured center oval in column 2 of the
clustering reagent plate.
Click “Close workspaces” for the cluster generator drawer to close shut. Select “Confirm” to confirm the tasks
listed in the software screen were completed.
DO NOT push or manually force reagent drawer to close. Do not place your
hands or fingers in or between the reagent drawer and cluster generator once
the automated opening or closure of the drawer has been started.
Select “Continue” on the software to initiate the 30-minute post-run wash.
Following the wash, select “Finish” on the completed screen to bring the software back to the home screen.
When the post-run wash is complete, the user will be allowed to access and dispose of the clustering reagent
plate, if necessary.