SECTION 5 REMOTE OPERATION
91
5.7.1 EXAMPLE OF STORING A PROGRAM
Step A. Address the UPC as a listener
:PROGram:NAME 1 <eos>
Step B. Select Program 1
NOTE: <eos> is a LINE FEED (0Ahex, 10dec)
Step C. Define program 1. Program 1 consists of steady-state and
optional transient segment parameter names and values. This is one
continuous string of commands and parameters
NOTE: Only commas may separate the values, NO Carriage
Returns or Line Feeds may separate the values as transmitted
:PROGram:DEFine
(Note: CURR:PROT:LEV and
CURR:PROT:TOUT are only
required if IPROT:STAT = ON)
FORM,3,
COUPLing,DIRECT,
XFMRRATIO,1.0,
FREQuency,60,
VOLTage1,115,
VOLTage2,115,
VOLTage3,115,
CURRent:LIMit,10
CURRent:PROTect:LEVel,50
CURRent:PROTect:TOUT,1
PHASe2,120,
PHASe3,240,
WAVEFORM1,1,
WAVEFORM2,1,
WAVEFORM3,1,
AUTORMS,1,
EVENTS,0,
OPTIONAL TRANSIENT PARAMETERS
SEGment,1,
FSEG,60,
VSEG1,120,
VSEG2,120,
VSEG3,120,
WFSEG1,1,
WFSEG2,1,
WFSEG3,1,
TSEG,.01,
LAST<eos>
Step D. Un-address UPC