5. Instrument installation
5.1. After unpacking, carefully check the packing list inside if the object is
complete and intact;
5.2. Determine whether the work environment to meet the foregoing
requirements, the ambient temperature 10 ~ 35 , relative humidity less than
85%, operating voltage 110-220V/50-60HZ.
5.3. The instrument is placed on a horizontal platform, the instrument should
avoid direct sunlight and be away from electromagnetic launchers and high-
power electrical devices, the use of the environment can’t have dust, corrosive
gases and vibration;
5.4. Around the instrument no obstacle to the flow of air around the instrument;
5.5. The company supplied with a power cord and make sure there are well-
grounded power outlet line;
5.6. Check the sample chamber, ensure that there is not any solution, and
foreign matter and the process of self-test to ensure that the sample
compartment lid is closed, you can’t half-open (this is very important otherwise
affect the instrument self-test results and normal use!).
5.7. Turn on the instrument. Then the instrument make self-test. After that, the
instrument can be operated normally. In case there is an error alarm halfway,
please refer to the chapter of instrument troubleshooting.
Note: Use only a grounded socket for the connection of this device to the power
supply. If you are not sure if the sockets are grounded, have this checked by a
qualified electrician. The power plug serves in addition to the power supply to
isolate the device quickly from the power source where necessary. During the
disconnection from the power source it must be made sure that the correct
power plug is pulled (for example by labeling the sockets). This is recommended
for long-term storage and can prevent potential dangers in the event of a fault.
Therefore make sure that the socket to which the device is connected is easy to
reach by each user at all times.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the PEAK INSTRUMENTS C-7200 Series and is the answer not in the manual?

PEAK INSTRUMENTS C-7200 Series Specifications

General IconGeneral
BrandPEAK INSTRUMENTS
ModelC-7200 Series
CategoryMeasuring Instruments
LanguageEnglish