Command Reference 8-53
+
The array size for :MEASure and :CONFigure, and the channels, are expression
data that must be within parentheses ( ).
The question mark at the end of the commands is only valid for :MEASure,
where a result is expected. It must be deleted for :CONFigure, which is a set
-
ting command only.
The default channels, which the counter use when you omit the channels in the
command, are printed in italics in the channel list on the following pages.
+
If you want to check what function and channels the counter is currently using,
send :CONF?
This query gives the same answer as :FUNC? in the SENSe subsystem
Measurement Function
Set up the Instrument, Perform Measurement, and Read Data
:MEASure
[:SCALar]<Measuring Function>?
[<Parameters>][,(<Channels>)]
:ARRay<Measuring Function>?
8
(<Array Size>)[,<Parameters>][,(<Channels>)]
:MEMory?
8
[<N>]
:MEMory<N>?