Command Reference 8-99
:HF :ACQuisition [:STATe]CNT-90XL
8 <Boolean>
Input C Acquisition ON/OFF
Switches the automatic acquisition system ON or OFF. ON means Automatic Ac
-
quisition, OFF means Manual Acquisition. When the instrument is switched from
remote to local operation, Automatic Acquisition mode is entered, irrespective of
the previous remote setting.
Parameters:
<Boolean> = «1 | ON» | «0 | OFF»
Returned format: 1|0¿
*RST condition: ON
:HF :FREQuency :CENTerCNT-90XL
8 <Numeric value>
Center Frequency
Sets the center frequency value for the RF input and is used when Manual Acquisi-
tion has been selected.
Parameters: <Numeric value> = a number between 3
*
10
8
(Hz) and 27
*
10
9
,40
*
10
9
,
46
*
10
9
or 60
*
10
9
(Hz), depending on the model number -27G, -40G, -46G or -60G
respectively.
Returned format: <Numeric value>¿
*RST condition: 300 MHz