EasyManua.ls Logo

Pendulum CNT-90 - Tinterval:AUTO; Tierror:Frequency

Pendulum CNT-90
264 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command Reference 8-101
:TINTerval :AUTO
8 «Boolean»
Smart Time Interval
By means of 4 time stamps (2 on each channel), the counter can determine which
event precedes the other. Thus you don’t have to set aside Input A as the start
channel.
:TIError :FREQuencyCNT-91
8 <Numeric value>
Set Basic TIE Frequency
An arbitrary frequency in the range 1 Hz to 100 MHz can be entered (increment =
1 Hz). Subsequent TIE measurements are made by continuous timestamping of
the input signal and the internal/external timebase clock. Observations of Wander,
for instance, can easily be made by means of this command and the function
:MEASure:ARRay:TIError? in conjunction with the built-in statistics/graphics fa
-
cilities.
Parameter: <Numeric value> = a number between 1 and 100
*
10
6
Hz in 1 Hz increments.