EasyManua.ls Logo

Pendulum CNT-90 - Pud; Rcl

Pendulum CNT-90
264 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command Reference 8-137
*PUD
8 Arbitrary block program data>
Protected User Data
Protected user data. This is a data area in which the user may write any data up to
64 characters. The data can always be read, but you can only write data after un
-
protecting the data area. A typical use would be to hold calibration information, us
-
age time, inventory control numbers, etc.
The content at delivery is: #234 FACTORY CALIBRATED ON: 19YY-MM-DD
YY = year, MM = month, DD = day
Returned format: <Arbitrary block response data>¿
Where:
<arbitrary block program data> is the data last programmed with *PUD.
Example
Send ®
:SYST:UNPR; *PUD 8 #240Calibrated 8 1993-07-16, 8 inven
-
tory 8 No.1234
# means that <arbitrary block program data> will follow.
2 means that the two following digits will specify the length of the data block.
40 is the number of characters in this example.
Complies with standards: IEEE 488.2 1987.
*RCL
8<Decimal data>
Recall
Recalls one of the up to 20 previously stored complete instrument settings from the
internal nonvolatile memory of the instrument.
Memory number 0 contains the power-off settings.
Parameters:
<Decimal data> = a number between 0 and 19.
Example:
SEND®
*RCL 8 10¿
Complies with standards: IEEE 488.2 1987.

Table of Contents

Other manuals for Pendulum CNT-90

Related product manuals