Command Reference 8-65
:MEASure [:VOLT] :MAXimum?
[ 8 («@1|@2»)]
Positive Peak Voltage
This command measures the positive peak voltage with the input DC coupled.
Parameters:
(«@1|@2») is the measurement channel
(@1) means input A
(@2) means input B
Complies with standards: SCPI 1991.0, confirmed.
:MEASure [:VOLT] :MINimum?
[ 8 («@1|@2»)]
Negative Peak Voltage
This command measures the negative peak voltage with the input DC coupled
Parameters:
(«@1|@2») is the measurement channel
(@1) means input A
(@2) means input B
Complies with standards: SCPI 1991.0, confirmed.