8-68 Command Reference
:MEASure :PERiod?
[8 [<expected value> [,<resolution>]][,<(@«1|2|3»)>]]
Period
A traditional period time measurement is performed on a single period. Measuring
time set by the :ACQ:APER command does not affect the measurement.
The <expected value> and <resolution> are used to calculate the Measurement
Time ([:SENSe]:ACQuisition:APERture).
Parameters:
<expected value> is the expected Period,
<resolution> is the required resolution,
<(@«1|2|3»)> is the measurement channel:
(@1) means input A
(@2) means input B
(@3) means input C (RF input option)
If you omit the channel, the instrument measures on input A (@1).
Complies with standards: SCPI 1991.0, confirmed.
:MEASure :PERiod :AVERage?
[8 [<expected value> [,<resolution>]][,<(@«1|2|3»)>]]
Period
A traditional period time measurement is performed on multiple periods. Measuring
time set by the :ACQ:APER command determines the resolution.
The <expected value> and <resolution> are used to calculate the Measurement
Time ([:SENSe]:ACQuisition:APERture).
Parameters:
<expected value> is the expected Period,
<resolution> is the required resolution,
<(@«1|2|3»)> is the measurement channel:
(@1) means input A
(@2) means input B
(@3) means input C (RF input option)
If you omit the channel, the instrument measures on input A (@1).
Complies with standards: SCPI 1991.0, confirmed.