8-94 Command Reference
:FREQuency :BURSt :PREScaler [:STATe]
8 <Boolean>
Prescaler ¸2 on Input A & Input B
The burst frequency limit is 300 MHz if the prescaler is ON and 160 MHz if it is
OFF.
Parameters: <Boolean> = (1/ON | 0/OFF)
Returned format: 1|0¿
*RST condition: ON
:FREQuency :BURSt :APERture
8 «<Numeric value>|MIN|MAX»
Burst Measuring Time
Sets the time length within a burst during which the burst frequency is measured.
Parameters: <Numeric value> = a number between 20 ns and 2 s.
Returned format: <Numeric value>¿
*RST condition: 200 ms