Command Reference 8-81
:MEMory :DATA :RECord :SAVE
8 <Dataset Number>[,<Label>]
Save Measurement Data Array to Internal FLASH Memory
One of the eight (0-7) memory positions must be entered, but you can also enter
an optional name (max 6 characters) for easier recognition.
A default name will be assigned automatically if you omit the <Label> parameter. It
represents the abbreviated measurement function and the channel. For example:
Period Single A will read PerA.
If the instrument is not in Hold when this command is sent, then Execution Error
(-200) will be placed in the error queue.
If the instrument is not in Statistics Mode when this command is sent, then Settings
Conflict Error (-221) will be placed in the error queue.
If specified <Dataset> already contains data, then Directory Full Error (-255) will be
placed in the error queue.
If there are more than 32000 samples to save, only the last 32000 will be saved
without notification to the operator.
:MEMory :DATA :RECord :SETTings?
8 <Dataset Number>
Recall Instrument Settings Used for Specified Dataset
The command returns the instrument settings used when the specified <Dataset>
was saved. The format is the same as for :SYSTem:SET.