EasyManua.ls Logo

Pendulum CNT-90 - Measure:Frequency:Ratio; Measure [:VOLT]:Ncycles

Pendulum CNT-90
264 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command Reference 8-63
:MEASure :FREQuency :RATio?
[8 [<expected value> [,<resolution>]][,<(@«1|2|3»)>,<(@«1|2|3»)>]]
Frequency Ratio
Frequency ratio measurements between two inputs.
Example:
SEND®
:MEAS:FREQ:RAT? 8 (@1),(@3)
READ¬ 2.345625764333E+000
This example measures the ratio between input A and input C.
+
The channel is expression data and must be within parentheses ( ).
Parameters: <expected value> and <resolution> are ignored
<(@«1|2|3»)>,<(@«1|2|3»)> are the measurement channels:
(@1) means input A
(@2) means input B
(@3) means input C (RF input option)
If you omit the channels, the instrument measures between input A and input B.
Complies with standards: SCPI 1991.0, confirmed.
:MEASure [:VOLT] :NCYCles?
[8 [(@1|@2|@3)]
Number of Cycles in Burst
If :FREQ:BURSt is active, this function measures the number of cycles in each
burst.
Returned format: <Numeric value (integer)>
Example:
SEND® :MEAS:NCYC? (@3)
READ¬
2356¿
This example shows a measurement on the RF channel.
+
The channel is expression data and must be within parentheses ( ).
<(@«1|2|3»)>,<(@«1|2|3»)> are the measurement channels:
(@1) means input A
(@2) means input B
(@3) means input C (RF input option)

Table of Contents

Other manuals for Pendulum CNT-90

Related product manuals